undefined
+
  • undefined

Expansion method vacuum device

This vacuum standard device is mainly used for verification/calibration of capacitance film vacuum gauges and thermocouple vacuum gauges in the range of (1.3 × 105~1.3 × 10-1) Pa. It mainly has the following characteristics: The 1.1 takes into account both the comparative method and the expansion method, which is convenient for customers to master flexibly and ensure that the equipment can be used for a long time; All instruments and equipment selected as part of the 1.2 expansion process provide a long term warranty. The standard capacitor film gauge for comparison method is guaranteed by the original factory. 1.3 using special software for data acquisition and processing, high efficiency; The 1.4 system software is upgraded free of charge for life, and the overall level of the device has reached the domestic advanced level.

Key words: standard leak | MPCVD | capacitance film gauge vacuum gauge | high and low vacuum calibration device | helium leak calibration device

Telephone:

Product Details

I. Overview
This vacuum standard device is mainly used for verification/calibration of capacitance film vacuum gauges and thermocouple vacuum gauges in the range of (1.3 × 105~1.3 × 10-1) Pa. It mainly has the following characteristics:
The 1.1 takes into account both the comparative method and the expansion method, which is convenient for customers to master flexibly and ensure that the equipment can be used for a long time;
All instruments and equipment selected as part of the 1.2 expansion process provide a long term warranty. The standard capacitor film gauge for comparison method is guaranteed by the original factory.
1.3 using special software for data acquisition and processing, high efficiency;
The 1.4 system software is upgraded free of charge for life, and the overall level of the device has reached the domestic advanced level.




Two main technical indicators
2.1 equipment: expansion system, air extraction system, pre-pressure measurement system, control system and other components.
Principle of 2.2 technology: static comparison method, primary expansion;
2.3 measurement range:(1.3 × 105~1.332 × 10 ) Pa;
2.4 ultimate vacuum degree: ≤ 1 × 10-4Pa;
Technical requirements for 2.5 uncertainty::
(1.332 × 105~1.332Pa × 103)Pa U ≤ 0.08 % k = 2 static comparison method;
(1.332 × 103~13.32Pa) U≤ 0.2% k = 2 primary expansion;
(13.32 × 10~1.332Pa) U≤ 0.5% k = 2 primary expansion;
(1.332 × 10~0.1332Pa) U≤ 1% k = 2 primary expansion








Related Products

Product inquiry

If you are interested in our products, please fill in the relevant information on this page and submit it, and there will be staff to contact you in time.
Submit Message